Draeger Medical Systems; Inc.   Apr-10-2014       Draeger Oxylog 3000 plus ventilators – Draeger Oxylog 3000 plus ventilators; emergency and transport ventilator. Part number: 5704833         The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation; gives an audible alarm and displays the error message Poti unplugged. Del Mar Reynolds Medical; Ltd. Apr-12-2014       Arkon Anesthesia Delivery System – SpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed.  This system includes a ventilator.    The System may go into a controlled failed state if a series of parameters on the user interface are activated in a specific sequence. Newport Medical Instruments Inc            Apr-14-2014       Ventilator – Newport HT70  and HT70 Plus Ventilators; Model Number: HT70 and HT70 Plus.Product Usage:The HT70 ventilator system is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically; the HT70 family of ventilators is applicable for in...